Products  >  Laboratory Instruments  >  Chemystry Analyzer
 

Overall performance
Equipment type          Fully automatic, STAT priority
Analysis speed             Constant speed 300Test/Hour and 500 Test/Hour with ISE
Test principle               Colorimetry, turbidimetry
Analysis method         End-point, 2 kinetics, fixed-time, turbidimeter.
Analysis method         Support single/double wavelength, linear and non-linear calibration
Sample System
Sample Tray                60 sample positions, random access of all samples (STAT priority)
Sample Volume          2-50ul, 0.1ul (each step)
Sample probe             Liquid level detection, the remaining detection, collision protection etc.
Sample Containers    Sample cups, Micro tube, Original blood-collection tube, etc.
Needle Cleaning:       Self-wash after every step, it can reduce the across-contaminated
Washing System:        8 needles with 4 pumps. Self-washinig in time with test
Reagent System
Reagent Tray              45 reagent positions. Open reagents system
Reagent Volume        R1:25-300ul, R2:10-150ul, 0.1ul increment.
Reagent Probe           Liquid level detection, the remaining detection, collision protection etc.
Reagent Refrigeration    4-15°C, independent power supply, 24-hour non-stop refrigeration
Reaction System
Reaction Tray              Rotating tray, 120 cuvettes
Cuvette                        Reusable high quality cuvette, Optical diameter 6mm
Reaction Volume       180-500ul
Incubation system     Bi-directional liquid-air device
Temperature              37°C ± 0.1°C with real-time display. Refrigeration for reagent tray (2-8°C)
Reaction Time:           About 10 minutes
Mixing Needle:          Self-mixing the reaction liquid
Liquid Sensor:            Self-detecting of the liquid level, tracing the rest volume
Optical System
Light source                12V, 35W halogen lamps, over 2000 hours
Photoelectron road    after spectrophotometry
Wavelength                 340,405,450,510,546,578,630,700nm
Absorbance range      0-3.500Abs
Resolution                    0.0001Abs
Operation System
PC operating system          Windows XP/win7 or above
Analysis control software   English version graphical operating software
Report printing                     Report formats support the user-defined mode,
Analysis method                  QC and state information etc.
System connection               RS-232
Working Conditions
Power supply                      AC 220V±22V, 50Hz±1Hz, power 1.5KVA
Ambient Temperature      15-30 ℃
Humidity                              35-80%
Water consumption          7L/Hour
Dimension                           Bench top: 750mm*520mm*600mm

 
 
 
 
 

Copyright © 2017 Easy Access International Co., Ltd. All Rights Reserved